ELECTRICAL

Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Equipment to Output
DC VoltageF
0 mV to 330 mV 0.002 % of reading + 1 µV Agilent 5520A
0 V to 3.3 V 0.001 1 % of reading + 2 µV
0 V to 33 V 0.001 2 % of reading + 20 µV
30 V to 330 V 0.001 8 % of reading + 150 µV
100 V to 1 000 V 0.001 5 % of reading + 1 500 µV
Equipment to Measure
DC VoltageF
0 mV to 100 mV 0.000 9 % of reading + 0.3 µV Agilent 3458A
100 mV to 1 V 0.000 8 % of reading + 0.3 µV
1 V to 10 V 0.000 8 % of reading + 0.5 µV
10 V to 100 V 0.001 % of reading + 30 µV
100 V to 1 000 V 0.001 % of reading + 0.1 mV
Equipment to Output
DC CurrentF
33 µA to 330 µA 0.015 % of reading + 0.02 µA Fluke 5520A
0.33 mA to 3.3 mA 0.010 % of reading + 0.05 µA
3.3 mA to 33 mA 0.010 % of reading + 0.25 µA
33 mA to 330 mA 0.010 % of reading + 2.5 µA
0.33 A to 1.1 A 0.010 % of reading + 40 µA
1.1 A to 2.99 999 A 0.038 % of reading + 40 µA
1.1 A to 11 A 0.050 % of reading + 500 µA
11 A to 20.5 A 0.10 % of reading + 750 µA
20.5 A to 150 A 0.5 % of output + 0.14 A Fluke 5520A / 50 turn coil
150 A to 1 025 A 0.5 % of output + 0.5 A
Equipment to Measure
DC CurrentF
0 nA to 100 nA 0.000 2 % of reading + 5 pA Agilent 3458A
100 nA to 1 µA 0.000 2 % of reading + 5 pA
1 µA to 10 µA 0.000 2 % of reading + 10 pA
10 µA to 100 µA 0.000 2 % of reading + 100 pA
100 µA to 1 mA 0.000 2 % of reading + 1 nA
1 mA to 10 mA 0.000 2 % of reading + 10 nA
10 mA to 100 mA 0.000 2 % of reading + 100 nA
100 mA to 1 A 0.000 2 % of reading + 2 µA
1 A to 3 A 0.12 % of reading + 0.6 mA Agilent 34401A
3 A to 6 A 0.2 % of reading + 4 mA Fluke 87V
6 A to 10 A 0.2 % of reading + 2 mA Fluke 87V


















 

Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Equipment to Output
AC Voltage
At the listed FrequenciesF
Fluke 5520A
10 Hz to 45 kHz 1 mV to 32.999 mV 0.08 % of reading + 6 µV
45 Hz to 10 kHz 1 mV to 32.999 mV 0.015 % of reading + 6 µV
10 kHz to 20 kHz 1 mV to 32.999 mV 0.02 % of reading + 6 µV
20 kHz to 50 kHz 1 mV to 32.999 mV 0.1 % of reading + 6 µV
50 kHz to 100 kHz 1 mV to 32.999 mV 0.35 % of reading + 12 µV
100 kHz to 500 kHz 1 mV to 32.999 mV 0.8 % of reading + 50 µV
Equipment to Output
AC Voltage
At the listed FrequenciesF
10 Hz to 45 Hz 33 mV to 329.999 mV 0.03 % of reading + 8 µV
10 Hz to 45 Hz 33 mV to 329.999 mV 0.015 % of reading + 8 µV
45 Hz to 10 kHz 33 mV to 329.999 mV 0.016 % of reading + 8 µV
10 kHz to 20 kHz 33 mV to 329.999 mV 0.035 % of reading + 8 µV
20 kHz to 50 kHz 33 mV to 329.999 mV 0.08 % of reading + 32 µV
50 kHz to 100 kHz 33 mV to 329.999 mV 0.2 % of reading + 70 µV
Equipment to Output
AC Voltage
At the listed FrequenciesF
10 Hz to 45 Hz 0.33 V to 3.299 99 V 0.03 % of reading + 50 µV
10 Hz to 45 Hz 0.33 V to 3.299 99 V 0.015 % of reading + 60 µV
45 Hz to 10 kHz 0.33 V to 3.299 99 V 0.019 % of reading + 60 µV
10 kHz to 20 kHz 0.33 V to 3.299 99 V 0.03 % of reading + 50 µV
20 kHz to 50 kHz 0.33 V to 3.299 99 mV 0.07 % of reading + 125 µV
50 kHz to 100 kHz 0.33 V to 3.299 99 mV 0.24 % of reading + 600 µV
Equipment to Output
AC Voltage
At the listed FrequenciesF
10 Hz to 45 Hz 3.3 V to 32.999 9 V 0.03 % of reading + 650 µV
10 Hz to 45 Hz 3.3 V to 32.999 9 V 0.015 % of reading + 600 µV
45 Hz to 10 kHz 3.3 V to 32.999 9 V 0.024 % of reading + 600 µV
10 kHz to 20 kHz 3.3 V to 32.999 9 V 0.035 % of reading + 600 µV
20 kHz to 50 kHz 3.3 V to 32.999 9 mV 0.09 % of reading + 1 600 µV















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Equipment to Output
AC Voltage
At the listed FrequenciesF
Fluke 5520A
45 Hz to 1k Hz 33 V to 329.999 V 0.019 % of reading + 2 000 µV
45 Hz to 1k Hz 33 V to 329.999 V 0.02 % of reading + 6 000 µV
1k Hz to 10 k Hz 33 V to 329.999 V 0.025 % of reading + 6 000 µV
10 kHz to 20 kHz 33 V to 329.999 V 0.03 % of reading + 6 000 µV
20 kHz to 50 kHz 33 V to 329.999 V 0.2 % of reading + 50 000 µV
Equipment to Output
AC Voltage
At the listed FrequenciesF
45 Hz to 1k Hz 330 V to 1 020 V 0.03 % of reading + 10 000 µV
45 Hz to 1k Hz 330 V to 1 020 V 0.03 % of reading + 10 000 µV
1k Hz to 10 k Hz 330 V to 1 020 V 0.3 % of reading + 10 000 µV
Equipment to Measure
AC Voltage
At the listed Frequencies
ACBAND < 2 MHzF
Agilent 3458A
1 Hz to 40 Hz 0 mV to 10 mV 0.03 % of reading + 0.03 mV
40 Hz to 1 kHz 0 mV to 10 mV 0.02 % of reading + 0.011 mV
1 kHz to 20 k Hz 0 mV to 10 mV 0.03 % of reading + 0.011 mV
20 kHz to 50 kHz 0 mV to 10 mV 0.1 % of reading + 0.011 mV
50 kHz to 100 kHz 0 mV to 10 mV 0.5 % of reading + 0.011 mV
100 kHz to 300 kHz 0 mV to 10 mV 4 % of reading + 0.02 mV
Equipment to Measure
AC Voltage
At the listed Frequencies
ACBAND < 2 MHzF
1 Hz to 40 Hz 100 mV to 10 V 0.007 % of reading + 0.004 V
40 Hz to 1kHz 100 mV to 10 V 0.007 % of reading + 0.002 V
1 kHz to 20 k Hz 100 mV to 10 V 0.014 % of reading + 0.002 V
20 kHz to 50 kHz 100 mV to 10 V 0.03 % of reading + 0.002 V
50 kHz to 100 kHz 100 mV to 10 V 0.08 % of reading + 0.002 V
100 kHz to 300 kHz 100 mV to 10 V 0.3 % of reading + 0.01 V
300 kHz to 1 MHz 100 mV to 10 V 1 % of reading + 0.01 V
1 MHz to 2 MHz 100 mV to 10 V 1.5 % of reading + 0.01 V















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Equipment to Measure
AC Voltage
At the listed Frequencies
ACBAND < 2 MHzF
Agilent 3458A
1 Hz to 40 Hz 33 V to 329.999 V 0.019 % of reading + 2 000 µV
45 Hz to 1k Hz 10 V to 100 V 0.02 % of reading + 0.04 V
40 Hz to 1kHz 10 V to 100 V 0.02 % of reading + 0.02 V
1 kHz to 20 k Hz 10 V to 100 V 0.02 % of reading + 0.02 V
20 kHz to 50 kHz 10 V to 100 V 0.35 % of reading + 0.02 V
50 kHz to 100 kHz 10 V to 100 V 0.12 % of reading + 0.02 V
100 kHz to 300 kHz 10 V to 100 V 0.4 % of reading + 0.1 V
300 kHz to 1 MHz 10 V to 100 V 1.5 % of reading + 0.1 V
Equipment to Measure
AC Voltage
At the listed Frequencies
ACBAND < 2 MHzF
1 Hz to 40 Hz 100 V to 1 000 V 0.04 % of reading + 0.4 V
40 Hz to 1kHz 100 V to 1 000 V 0.04 % of reading + 0.2 V
1 kHz to 20 k Hz 100 V to 1 000 V 0.06 % of reading + 0.2 V
20 kHz to 50 kHz 100 V to 1 000 V 0.12 % of reading + 0.2 V
50 kHz to 100 kHz 100 V to 1 000 V 0.3 % of reading + 0.2 V
Equipment to Measure
AC Voltage
At the listed Frequencies
ACBAND > 2 MHzF
45 Hz to 100 kHz 0 mV to 10 mV 0.09 % of reading + 0.06 mV
100 kHz to 1MHz 0 mV to 10 mV 1.2 % of reading + 0.05 mV
1 MHz to 4MHz 0 mV to 10 mV 7 % of reading + 0.07 mV
4 MHz to 8 MHz 0 mV to 10 mV 20 % of reading + 0.08 mV
Equipment to Measure
AC Voltage
At the listed Frequencies
ACBAND > 2 MHzF
45 Hz to 100 kHz 100 mV to 10 V 0.09 % of reading + 0.06 V
100 kHz to 1MHz 100 mV to 10 V 2 % of reading + 0.05 V
1 MHz to 4 MHz 100 mV to 10 V 4 % of reading + 0.07 V
4 MHz to 8 MHz 100 mV to 10 V 4 % of reading + 0.08 V
8 MHz to 10 MHz 100 mV to 10 V 15 % of reading + 0.01 V















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Equipment to Measure
AC Voltage
At the listed Frequencies
ACBAND > 2 MHzF
Agilent 3458A
45 Hz to 100 kHz 10 V to 100 V 0.12 % of reading + 0.02 V
45 Hz to 100 kHz 100 V to 1 000 V 0.3 % of reading + 1 V
Equipment to Output
AC Current
At the listed FrequenciesF
Fluke 5520A
10 Hz to 20 Hz 29 µA to 329.99 µA 0.2 % of reading + 0.1 µA
20 Hz to 45 Hz 29 µA to 329.99 µA 0.15 % of reading + 0.1 µA
45 Hz to 1 kHz 29 µA to 329.99 µA 0.13 % of reading + 0.1 µA
1 kHz to 5 kHz 29 µA to 329.99 µA 0.3 % of reading + 0.015 µA
5 kHz to 10 kHz 29 µA to 329.99 µA 0.8 % of reading + 0.2 µA
10 kHz to 30 kHz 29 µA to 329.99 µA 1.6 % of reading + 0.4 µA
Equipment to Output
AC Current
At the listed FrequenciesF
10 Hz to 20 Hz 0.33 mA to 3.299 9 mA 0.2 % of reading + 0.15 µA
20 Hz to 45 Hz 0.33 mA to 3.299 9 mA 0.13 % of reading + 0.15 µA
45 Hz to 1 kHz 0.33 mA to 3.299 9 mA 0.1 % of reading + 0.15 µA
1 kHz to 5 kHz 0.33 mA to 3.299 9 mA 0.2 % of reading + 0.2 µA
5 kHz to 10 kHz 0.33 mA to 3.299 9 mA 0.5 % of reading + 0.3 µA
10 kHz to 30 kHz 0.33 mA to 3.299 9 mA 1 % of reading + 0.6 µA
Equipment to Output
AC Current
At the listed FrequenciesF
10 Hz to 20 kHz 3.3 mA to 32.999 mA 0.18 % of reading + 2 µA
20 Hz to 45 Hz 3.3 mA to 32.999 mA 0.09 % of reading + 2 µA
45 Hz to 1 kHz 3.3 mA to 32.999 mA 0.04 % of reading + 2 µA
1 kHz to 5 kHz 3.3 mA to 32.999 mA 0.08 % of reading + 2 µA
5 kHz to 10 kHz 3.3 mA to 32.999 mA 0.2 % of reading + 3 µA
10 kHz to 30 kHz 3.3 mA to 32.999 mA 0.4 % of reading + 4 µA















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Equipment to Output
AC Current
At the listed FrequenciesF
Fluke 5520A
10 Hz to 20 Hz 33 mA to 329.99 mA 0.18 % of reading + 20 µA
20 Hz to 45 Hz 33 mA to 329.99 mA 0.09 % of reading + 20 µA
45 Hz to 1 kHz 33 mA to 329.99 mA 0.04 % of reading + 20 µA
1 kHz to 5 kHz 33 mA to 329.99 mA 0.10 % of reading + 50 µA
5 kHz to 10 kHz 33 mA to 329.99 mA 0.2 % of reading + 100 µA
10 kHz to 30 kHz 33 mA to 329.99 mA 0.4 % of reading + 200 µA
Equipment to Output
AC Current
At the listed FrequenciesF
10 Hz to 45 Hz 0.33 A to 1.099 99 A 0.18 % of reading + 100 µA
45 Hz to 1 kHz 0.33 A to 1.099 99 A 0.05 % of reading + 100 µA
1 kHz to 5 kHz 0.33 A to 1.099 99 A 0.6 % of reading + 1 000 µA
5 kHz to 10 kHz 0.33 A to 1.099 99 A 2.5 % of reading + 5 000 µA
Equipment to Output
AC Current
At the listed FrequenciesF
10 Hz to 45 Hz 1.1 A to 2.299 99 A 0.18 % of reading + 100 µA
45 Hz to 1 kHz 1.1 A to 2.299 99 A 0.06 % of reading + 100 µA
1 kHz to 5 kHz 1.1 A to 2.299 99 A 0.6 % of reading + 1 000 µA
5 kHz to 10 kHz 1.1 A to 2.299 99 A 2.5 % of reading + 5 000 µA
Equipment to Output
AC Current
At the listed FrequenciesF
45 Hz to 100 Hz 3 A to 10.999 9 A 0.06 % of reading + 2 000 µA
100 Hz to 1 kHz 3 A to 10.999 9 A 0.10 % of reading + 2 000 µA
1 kHz to 5 kHz 3 A to 10.999 9 A 3 % of reading + 2 000 µA
Equipment to Output
AC Current
At the listed frequenciesF
45 Hz to 100 Hz 11 A to 20.5 A 0.12 % of reading + 5 000 µA
100 Hz to 1 kHz 11 A to 20.5 A 0.15 % of reading + 5 000 µA
1 kHz to 5 kHz 11 A to 20.5 A 3 % of reading + 5 000 µA















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Equipment to Output
AC Current
At the listed FrequenciesF
Fluke 5520A / 50 turn coil
45 Hz to 65 Hz 20.5 A to 150 A 0.56 % of reading + 0.25 A
65 Hz to 440 Hz 20.5 A to 150 A 1 % of reading + 0.25 A
Equipment to Output
AC Current
At the listed FrequenciesF
45 Hz to 65 Hz 150 A to 1 025 A 0.56 % of reading + 0.9 A
65 Hz to 440 Hz 150 A to 1 025 A 1 % of reading + 0.9 A
Equipment to Output
AC Current
At the listed FrequenciesF
Agilent 3458A
10 Hz to 20 Hz 0 µA to 100 µA 0.4 % of reading + 0.03 µA
20 Hz to 45 Hz 0 µA to 100 µA 0.15 % of reading + 0.03 µA
45 Hz to 100 Hz 0 µA to 100 µA 0.06 % of reading + 0.03 µA
100 Hz to 5 kHz 0 µA to 100 µA 0.06 % of reading + 0.03 µA
Equipment to Output
AC Current
At the listed FrequenciesF
10 Hz to 20 Hz 1 mA to 100 mA 0.4 % of reading + 0.02 mA
20 Hz to 45 Hz 1 mA to 100 mA 0.15 % of reading + 0.02 mA
45 Hz to 100 Hz 1 mA to 100 mA 0.06 % of reading + 0.02 mA
100 Hz to 5 kHz 1 mA to 100 mA 0.03 % of reading + 0.02 mA
5 kHz to 20 kHz 1 mA to 100 mA 0.06 % of reading + 0.02 mA
20 kHz to 50 kHz 1 mA to 100 mA 0.4 % of reading + 0.04 mA
50 kHz to 100 kHz 1 mA to 100 mA 0.55 % of reading + 0.15 mA
Equipment to Output
AC Current
At the listed FrequenciesF
10 Hz to 20 Hz 100 mA to 1 A 0.4 % of reading + 0.2 mA
20 Hz to 45 Hz 100 mA to 1 A 0.16 % of reading + 0.2 mA
45 Hz to 100 Hz 100 mA to 1 A 0.08 % of reading + 0.2 mA
100 Hz to 5 kHz 100 mA to 1 A 0.1 % of reading + 0.2 mA
5 kHz to 20 kHz 100 mA to 1 A 0.3 % of reading + 0.2 mA
20 kHz to 50 kHz 100 mA to 1 A 1 % of reading + 0.4 mA















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Equipment to Measure
AC Current
At the listed FrequenciesF
Agilent 34401A
3 Hz to 5 Hz 1 A to 3 A 1.10 % of reading + 0.001 8 A
5 Hz to 10 Hz 1 A to 3 A 0.35 % of reading + 0.001 8 A
10 Hz to 5 kHz 1 A to 3 A 0.15 % of reading + 0.001 8 A
Equipment to Measure
AC Current
At the listed FrequenciesF
Fluke 87V
45 Hz to 2 kHz 3 A to 6 A 1 % of reading + 0.002 A
45 Hz to 2 kHz 6 A to 10 A 1 % of reading + 0.02 A
Equipment to Output
ResistanceF
0 to 11 Ω 0.004 % of reading + 1 mΩ Fluke 5520A
11 Ω to 33 Ω 0.003 % of reading + 15 mΩ
33 Ω to 110 Ω 0.002 8 % of reading + 14 mΩ
110 Ω to 330 Ω 0.002 8 % of reading + 2 mΩ
330 Ω to 1.1 kΩ 0.002 8 % of reading + 2 mΩ
1.1 kΩ to 3.3 kΩ 0.002 8 % of reading + 20 mΩ
3.3 kΩ to 11 kΩ 0.002 8 % of reading + 20 mΩ
11 kΩ to 33 kΩ 0.003 2 % of reading + 200 mΩ
33 kΩ to 110 kΩ 0.0032 % of reading + 200 mΩ
110 kΩ to 330 kΩ 0.003 2 % of reading + 2 Ω
330 kΩ to 1.1 MΩ 0.003 2 % of reading + 2 Ω
1.1 MΩ to 3.3 MΩ 0.006 0 % of reading + 30 Ω
3.3 MΩ to 11 MΩ 0.013 % of reading + 50 Ω
11 MΩ to 33 MΩ 0.025 % of reading + 2 500 Ω
33 MΩ to 110 MΩ 0.05 % of reading + 3 000 Ω
110 MΩ to 330 MΩ 0.03 % of reading + 10 000 Ω
330 MΩ to 1.1 G&Omega 1.5 % of reading + 500 000 Ω
Equipment to Measure
ResistanceF
0 to 10 Ω 0.001 5 % of reading + 50 µΩ Agilent 3458A
0 Ω to 100 Ω 0.001 2 % of reading + 500 µΩ
0 Ω to 1 kΩ 0.001 % of reading + 500 µΩ
0 Ω to 10 kΩ 0.001 % of reading + 5 mΩ
0 Ω to 100 kΩ 0.001 % of reading + 50 mΩ
0 Ω to 1 MΩ 0.001 5 % of reading + 2 Ω
0 Ω to 10 MΩ 0.005 % of reading + 100 Ω
0 kΩ to 100 MΩ 0.05 % of reading + 1 KΩ
0 Ω to 1 GΩ 0.5 % of reading + 10 KΩ















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Equipment to Output
Capacitance
At the listed FrequenciesF
Fluke 5520A
10 Hz to 10 kHz 0.19 nF to 0.399 9 nF 0.5 % of reading + 0.01 nF
10 Hz to 10 kHz 0.4 nF to 1.099 9 nF 0.5 % of reading + 0.01 nF
10 Hz to 3 kHz 1.1 nF to 3.299 9 nF 0.5 % of reading + 0.01 nF
10 Hz to 1 kHz 3.3 nF to 10.999 9 nF 0.25 % of reading + 0.01 nF
10 Hz to 1 kHz 11 nF to 32.999 9 nF 0.25 % of reading + 0.1 nF
10 Hz to 1 kHz 33 nF to 109.999 nF 0.25 % of reading + 0.1 nF
10 Hz to 1 kHz 110 nF to 329.999 nF 0.25 % of reading + 0.3 nF
10 Hz to 600 kHz 0.33 µF to 1.099 99 µF 0.25 % of reading + 1 nF
10 Hz to 300 kHz 1.1 µF to 3.299 99 µF 0.25 % of reading + 3 nF
10 Hz to 150 Hz 3.3 µF to 10.999 9 µF 0.25 % of reading + 10 nF
10 Hz to 120 Hz 11 µF to 32.999 9 µF 0.40 % of reading + 30 nF
10 Hz to 80 Hz 33 µF to 109.999 µF 0.45 % of reading + 100 nF
0 Hz to 50 Hz 110 µF to 329.999 µF 0.45 % of reading + 300 nF
0 Hz to 20 Hz 0.33 µF to 1.099 99 mF 0.45 % of reading + 1 µF
0 Hz to 6 Hz 1.1 mF to 3.299 9 mF 0.45 % of reading + 3 µF
0 Hz to 2 Hz 3.3 mF to 10.999 9 mF 0.45 % of reading + 10 µF
0 Hz to 0.6 Hz 11 mF to 32.999 9 mF 0.75 % of reading + 30 µF
0 Hz to 0.2 Hz 33 mF to 110 mF 1.1 % of reading + 100 µF
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type BF
600 °C to 800 °C 0.44 °C Fluke 5520A / SC600
Electrical Simulation of
Thermocouple Output
800 °C to 1 000 °C 0.34 °C
1 000 °C to 1 550 °C 0.30 °C
1 550 °C to 1 820 °C 0.33 °C
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type CF
0 °C to 150 °C 0.30 °C
150 °C to 650 °C 0.26 °C
650 °C to 1 000 °C 0.31 °C
1 000 °C to 1 800 °C 0.50 °C
1 800 °C to 2 316 °C 0.84 °C
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type EF
-250 °C to -100 °C 0.50 °C
-100 °C to -25 °C 0.16 °C
-25 °C to 350 °C 0.14 °C
350 °C to 650 °C 0.16 °C
650 °C to 1 000 °C 0.21 °C















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type JF
-210 °C to -100 °C 0.27 °C Fluke 5520A / SC600
Electrical Simulation of
Thermocouple Output
-100 °C to -30 °C 0.16 °C
-30 °C to 150 °C 0.14 °C
150 °C to 760 °C 0.17 °C
760 °C to 1 200 °C 0.23 °C
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type KF
-200 °C to -100 °C 0.33 °C
-100 °C to -25 °C 0.18 °C
-25 °C to 120 °C 0.16 °C
120 °C to 1 000 °C 0.26 °C
1 000 °C to 1 372 °C 0.40 °C
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type LF
-200 °C to -100 °C 0.37 °C
-100 °C to 800 °C 0.26 °C
-800 °C to 900 °C 0.17 °C
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type NF
-200 °C to -100 °C 0.40 °C
-100 °C to -25 °C 0.22 °C
-25 °C to 120 °C 0.19 °C
120 °C to 410 °C 0.18 °C
410 °C to 1 300 °C 0.27 °C
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type RF
0 °C to 250 °C 0.57 °C
250 °C to 400 °C 0.35 °C
400 °C to 1 000 °C 0.33 °C
1 000 °C to 1 767 °C 0.40 °C
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type SF
0 °C to 250 °C 0.47 °C
250 °C to 1 000 °C 0.36 °C
1 000 °C to 1 400 °C 0.37 °C
1 400 °C to 1 767 °C 0.46 °C
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type TF
-250 °C to -150 °C 0.63 °C
-150 °C to 0 °C 0.24 °C
0 °C to 120 °C 0.16 °C
120 °C to 400 °C 0.14 °C
Temperature Calibration,
Indication and Control
Equipment used with
Thermocouple Type UF
-200 °C to 0 °C 0.56 °C
0 °C to 600 °C 0.27 °C















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Temperature Calibration,
Indication and Control
Equipment used with
RTD Pt 385, 100 ΩF
-200 °C to -80 °C 0.05 °C Fluke 5520A / SC600
Electrical Simulation of
Thermocouple Output
-80 °C to 0 °C 0.05 °C
0 °C to 100 °C 0.07 °C
100 °C to 300 °C 0.09 °C
300 °C to 400 °C 0.10 °C
400 °C to 630 °C 0.12 °C
630 °C to 800 °C 0.23 °C
Temperature Calibration,
Indication and Control
Equipment used with
RTD Pt 3926, 100 ΩF
-200 °C to -80 °C 0.05 °C
-80 °C to 0 °C 0.05 °C
0 °C to 100 °C 0.07 °C
100 °C to 300 °C 0.09 °C
300 °C to 400 °C 0.10 °C
400 °C to 630 °C 0.12 °C
Temperature Calibration,
Indication and Control
Equipment used with
RTD Pt 3916, 100 ΩF
-200 °C to -190 °C 0.25 °C
-190 °C to -80 °C 0.04 °C
-80 °C to 0 °C 0.05 °C
0 °C to 100 °C 0.06 °C
100 °C to 260 °C 0.07 °C
260 °C to 300 °C 0.08 °C
300 °C to 400 °C 0.09 °C
400 °C to 600 °C 0.10 °C
600 °C to 630 °C 0.23 °C
Temperature Calibration,
Indication and Control
Equipment used with
RTD Pt 385, 200 ΩF
-200 °C to -80 °C 0.04 °C
-80 °C to 0 °C 0.04 °C
0 °C to 100 °C 0.04 °C
100 °C to 260 °C 0.05 °C
260 °C to 300 °C 0.12 °C
300 °C to 400 °C 0.13 °C
400 °C to 600 °C 0.14 °C
600 °C to 630 °C 0.16 °C















 
Measured Instrument,
Quantity or Gauge.
Range or Nominal Device
Size as Appropriate.
Calibration and Measurement
Capability Expressed as an Uncertainty.
Calibration Equipement
and Reference Standards Used.
Temperature Calibration,
Indication and Control
Equipment used with
RTD Pt 385, 500 ΩF
-200 °C to -80 °C 0.05 °C Fluke 5520A / SC600
Electrical Simulation of
Thermocouple Output
-200 °C to -80 °C 0.04 °C
-80 °C to 0 °C 0.05 °C
0 °C to 100 °C 0.05 °C
100 °C to 260 °C 0.06 °C
260 °C to 300 °C 0.08 °C
300 °C to 400 °C 0.08 °C
400 °C to 600 °C 0.09 °C
600 °C to 630 °C 0.11 °C
Temperature Calibration,
Indication and Control
Equipment used with
RTD Pt 385, 1 000 ΩF
-200 °C to -80 °C 0.03 °C
-80 °C to 0 °C 0.03 °C
0 °C to 100 °C 0.04 °C
100 °C to 260 °C 0.05 °C
260 °C to 300 °C 0.06 °C
300 °C to 400 °C 0.07 °C
400 °C to 600 °C 0.07 °C
600 °C to 630 °C 0.23 °C
Temperature Calibration,
Indication and Control
Equipment used with
RTD PtNi 385, 120 ΩF
-80 °C to 0 °C 0.08 °C
0 °C to 100 °C 0.08 °C
100 °C to 260 °C 0.14 °C
Temperature Calibration,
Indication and Control
Equipment used with
RTD Cu 427, 10 ΩF
-100 °C to -260 °C 0.03 °C